HANNA HI98196-10 Multiparameter pH/ORP/DO/Pressure/ Temperature Waterproof Meter,10m cable
- Brand: Hanna Instruments
- Product Code: HI98196-10
- Availability: 10 - 14 Working Days
- Rp.28,950,000
Tags: HANNA, pH, Conductivity, TDS, Turbidity, Dissolved Oxygen, Tester
The HI7698196 is a multiparameter pH/EC/Temperature probe for use with the HI98196 portable meter. It features a quick connect that makes a waterproof connection with the meter. Sensors are automatically recognized by the probe and meter when connected. Any ports not used on the probe will not have the parameter displayed on the meter. Sensor replacement is quick and easy with field replaceable, screw type connectors that are color coded for easy sensor identification. The probe features a multistrand-multiconductor shielded cable with 4m, 10m, 20m, and 40m lengths available. It’s rugged, waterproof design makes it ideal for field use.
Includes:
- (1) HI7698196 probe
- (1) HI7698194-1 pH/ORP sensor
- (1) HI7698194-2 DO sensor
- (1) HI 76981942 Probe maintenance kit (HI 7042S DO electrolyte solution, DO membrane caps (5), O-rings for sensors (5), syringe with grease to lubricate the O-rings)
- (1) HI9298194 Hanna PC software
- (1) HI920015 micro USB cable
- (4) 1.5V AA batteries
- (1) Instruction manual
- (1) Quick start guide
- (1) Quality certificate
- (1) Rugged carrying case with custom thermoformed insert.
Specifications:
- pH Range: 0.00 to 14.00 pH
- pH Resolution: 0.01 pH
- pH Accuracy: ±0.02 pH
- pH Calibration: automatic one, two, or three points with automatic recognition of five standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01) or one custom buffer
- mV Range: ±600.0 mV
- mV Resolution: 0.1 mV
- mV Accuracy: ±0.5 mV
- ORP Range: ±2000 mV
- ORP Resolution: 0.1 mV
- ORP Accuracy: ±1.0 mV
- ORP Calibration: automatic at one custom point (relative mV)
- DO Range: 0.0 to 500.0%; 0.00 to 50.00 ppm (mg/L)
- DO Resolution: 0.1%; 0.01 ppm (mg/L)
- DO Accuracy: 0.0 to 300.0% saturation: ±1.5% of reading or ±1.0% saturation whichever is greater, 0.00 to 30.00 ppm (mg/L): ±1.5% of reading or ±0.10 ppm (mg/L), whichever is greater, 30.00 ppm (mg/L) to 50.00 ppm (mg/L): ±3% of reading, 300.0 to 500.0% saturation: ±3% of reading
- DO Calibration: automatic one or two point at 100% saturation (8.26 mg/L) and 0% saturation (0 mg/L).
- Atmospheric Pressure Range: 450 to 850 mm Hg; 17.72 to 33.46 in Hg; 600.0 to 1133.2 mbar; 8.702 to 16.436 psi; 0.5921 to 1.1184 atm; 60.00 to 113.32 kPa
- Atmospheric Pressure Resolution: 0.1 mm Hg; 0.01 in Hg; 0.1 mbar; 0.001 psi; 0.0001 atm; 0.01 kPa
- Atmospheric Pressure Accuracy: ±3 mm Hg within ±15°C from the temperature during calibration
- Atmospheric Pressure Calibration: automatic at one custom point
- Temperature Range: -5.00 to 55.00 °C, 23.00 to 131.00 °F, 268.15 to 328.15 K
- Temperature Resolution: 0.01 K, 0.01 °C, 0.01 °F
- Temperature Accuracy: ±0.15 °C; ±0.27 °F; ±0.15 K
- Temperature Calibration: automatic at one custom point
- Temperature Compensation: automatic from -5 to 55 °C (23 to 131 °F)
- Logging Memory: 45,000 records (continuous logging or log-on-demand of all parameters)
- Logging Interval: one second to three hours
- PC Connectivity: USB to PC with HI9298194 Hanna software installed
- Battery Type/Life: 1.5V AA batteries (4) / approximately 360 hours of continuous use without backlight (50 hours with backlight)
- Environment: 0 to 50°C (32 to 122°F); RH 100% (IP67)
- Dimensions: 185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4”)
- Weight: 400 g (14.2 oz.)